A versatile calibration procedure for portable coded aperture gamma cameras and RGB-D sensors - CEA - Commissariat à l’énergie atomique et aux énergies alternatives Accéder directement au contenu
Article Dans Une Revue Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment Année : 2018

A versatile calibration procedure for portable coded aperture gamma cameras and RGB-D sensors

Résumé

The present paper proposes a versatile procedure for the geometrical calibration of coded aperture gamma cameras and RGB-D depth sensors, using only one radioactive point source and a simple experimental set-up. Calibration data is then used for accurately aligning radiation images retrieved by means of the -camera with the respective depth images computed with the RGB-D sensor. The system resulting from such a combination is thus able to retrieve, automatically, the distance of radioactive hotspots by means of pixel-wise mapping between gamma and depth images. This procedure is of great interest for a wide number of applications, ranging from precise automatic estimation of the shape and distance of radioactive objects to Augmented Reality systems. Incidentally, the corresponding results validated the choice of a perspective design model for a coded aperture -camera.
Fichier non déposé

Dates et versions

hal-01707584 , version 1 (12-02-2018)

Identifiants

Citer

V. Paradiso, A. Crivellaro, K. Amgarou, N. Blanc de Lanaute, P. Fua, et al.. A versatile calibration procedure for portable coded aperture gamma cameras and RGB-D sensors. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018, 886, pp.125-133. ⟨10.1016/j.nima.2017.12.065⟩. ⟨hal-01707584⟩
141 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More