Americium isotope analysis by Thermal Ionization Mass Spectrometry using the Total Evaporation Method - CEA - Commissariat à l’énergie atomique et aux énergies alternatives Accéder directement au contenu
Article Dans Une Revue International Journal of Mass Spectrometry Année : 2018

Americium isotope analysis by Thermal Ionization Mass Spectrometry using the Total Evaporation Method

Résumé

In this paper, a simple analytical procedure was developed to measure americium isotope ratios and concentration with high accuracy. The method was tested during the participation in a round robin test organized by the Analytical Methods Committee of the French Atomic Energy Commission. The measurements were performed by Thermal Ionization Mass Spectrometry using the total evaporation method, which is a reference technique for determining actinide isotopic compositions. Expanded uncertainties were estimated at 0.1 % and 0.8 % for the 241 Am/ 243 Am isotope ratio and the Am concentration, respectively. Compared to the assigned value, biases below 0.0001% and 0.02% were calculated for the 241 Am/ 243 Am isotope ratio and the Am concentration, respectively.
Fichier principal
Vignette du fichier
Publi_Am.pdf (657.68 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

cea-03470762 , version 1 (08-12-2021)

Identifiants

Citer

Alexandre Quemet, Alexandre Ruas, Vincent Dalier, Cédric Rivier. Americium isotope analysis by Thermal Ionization Mass Spectrometry using the Total Evaporation Method. International Journal of Mass Spectrometry, 2018, 431, pp.8-14. ⟨10.1016/j.ijms.2018.05.017⟩. ⟨cea-03470762⟩

Collections

CEA DEN DEN-MARCO
54 Consultations
128 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More