In quest of the optimal coincidence resolving time in TDCR LSC

Résumé

This paper presents studies of the influence of the coincidence resolving time on the activity calculated by the Triple-to-Double Coincidences Ratio (TDCR) method in Liquid Scintillation (LS) counting. Recently, published methods for the correction for accidental coincidences in TDCR counting open the possibility to use resolving times up to several µs, long enough not to miss true coincidences and to study the effects of delayed fluorescence. 3H, 14C, 55Fe and 63Ni LS-sources in UltimaGold (UG), UG LLT and Toluene+PPO cocktails were measured using a TDCR counter connected to a digitizer working in list-mode. The necessary resolving time to include 99.9% of the logical sum of double (D) coincidences was found to be 1.2 µs for 3H, 1 µs for 55Fe and 500 ns for 63Ni in UG. The activity of all LS-sources was calculated using the TDCR method for resolving times from 10 ns to 2 µs and a significant dependence between the calculated activity and resolving time was observed. A dedicated Monte Carlo (MC) code was used to simulate list-mode data from TDCR measurements. The simulation results suggest that the 3H activity calculated by the TDCR method is overestimated regardless of the used resolving time if delayed fluorescence is present which is not described by the used ionization quenching function. Efficiency variation measurements of 3H in UG LLT show a strong dependence of the optimal kB parameter on the used resolving time: 85 µm/MeV at 40 ns and 110 µm/MeV at 200 ns, leading to 2.5% difference in calculated activity. In the framework of this study the efficiency variation methods by chemical quenching and gray filters were compared and a difference of 60 µm/MeV between the two was observed. The results from this article demonstrate that regardless of the available corrections for accidental coincidences, it is not advisable to increase the resolving time beyond what is necessary to register all prompt fluorescence events. Moreover, even for short coincidence resolving times, delayed fluorescence could have a significant influence on the activities calculated by the TDCR method.
Fichier principal
Vignette du fichier
article_ChavdarDutsov.pdf (607.42 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

cea-03085343 , version 1 (22-12-2020)

Identifiants

Citer

Chavdar Dutsov, Philippe Cassette, Krasimir Mitev, Benoît Sabot. In quest of the optimal coincidence resolving time in TDCR LSC. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2021, 987, pp.164846. ⟨10.1016/j.nima.2020.164846⟩. ⟨cea-03085343⟩
103 Consultations
121 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More