Skip to Main content Skip to Navigation
Journal articles

Dosimetry formalism and calibration procedure for electronic brachytherapy sources in terms of absorbed dose to water

Abstract : The LNE-LNHB has developed a methodology to standardize electronic brachytherapy sources in terms of absorbed dose to water. It is based on the measurement of the air-kerma rate at a given distance from the source and the Monte Carlo calculation of a conversion factor. This factor converts the air kerma in measurement conditions into absorbed dose to water at a 1 cm reference depth in a water phantom. As a first application, the method was used to calibrate a Zeiss INTRABEAM system equipped with its 4 cm diameter spherical applicator. The absorbed-dose rate value obtained in the current study was found significantly higher than that provided by the manufacturer in line with the observations already reported by a few other teams.
Complete list of metadatas

Cited literature [9 references]  Display  Hide  Download

https://hal-cea.archives-ouvertes.fr/cea-02611673
Contributor : Marie-France Robbe <>
Submitted on : Monday, May 18, 2020 - 3:45:27 PM
Last modification on : Saturday, October 10, 2020 - 3:07:35 AM

File

 Restricted access
To satisfy the distribution rights of the publisher, the document is embargoed until : 2022-05-18

Please log in to resquest access to the document

Identifiers

  • HAL Id : cea-02611673, version 1

Citation

Abdullah Abudra'A, Bruno Chauvenet, Jean Gouriou, Johann Plagnard, Itti Ramona, et al.. Dosimetry formalism and calibration procedure for electronic brachytherapy sources in terms of absorbed dose to water. Physics in Medicine & Biology, In press. ⟨cea-02611673⟩

Share

Metrics

Record views

91