Graphite monochromator for actinide L-line energy dispersive X-ray fluorescence analysis in liquid sample

Abstract : The concentrations of actinides U, Np, Pu, Am, and Cm are determined by L-line X-ray fluorescence measurement in a hot cell. A specific device was built to focalize only on the energy range from 13 to 15 keV by means of a graphite monochromator. The resulting spectrum is characteristic of the system, which has been developed, and requires the development of specific processing software for quantitative analysis. This paper proposes a new approach of modeling and correction of matrix effects for these specific spectra. This correction requires a precise chemical description of the media. Several results were compared with other analytical techniques and very good agreement was found.
Complete list of metadatas

Cited literature [3 references]  Display  Hide  Download

https://hal-cea.archives-ouvertes.fr/cea-02389398
Contributor : Amplexor Amplexor <>
Submitted on : Monday, December 2, 2019 - 3:01:12 PM
Last modification on : Tuesday, April 28, 2020 - 11:28:14 AM
Long-term archiving on: : Tuesday, March 3, 2020 - 7:52:14 PM

File

201600001072.pdf
Files produced by the author(s)

Identifiers

Collections

CEA | DEN

Citation

E. Esbelin. Graphite monochromator for actinide L-line energy dispersive X-ray fluorescence analysis in liquid sample. X-Ray Spectrometry, Wiley, 2014, 43 (4), pp.198-208. ⟨10.1002/xrs.2540⟩. ⟨cea-02389398⟩