Abstract : An original algorithm for real-time In Vivo Dosimetry (IVD) based on Radioluminescence (RL) of dosimetric-grade Al$_2$O$_3$:C crystals is described and demonstrated in reference conditions with 12-MV photon beams from a Saturne 43 linear accelerator (LINAC), simulating External Beam Radiation Therapy (EBRT) treatments. During the course of irradiation, a portion of electrons is trapped within the Al$_2$O$_3$:C crystal while another portion recombines and generates RL, recorded on-line using an optical fiber. The RL sensitivity is dose-dependent and increases in accordance with the concentration of trapped electrons. Once irradiation is completed, the Al$_2$O$_3$:C crystal is reset by laser light (reusable) and the resultant OSL (Optically Stimulated Luminescence) is also collected back by the remote RL-OSL reader and finally integrated to yield the absorbed dose. During irradiation, scintillation and Cerenkov lights generated within the optical fiber (“stem effect”) are removed by a time-discrimination method involving a discriminating unit and a fiber-coupled BGO scintillator placed in the irradiation room, next to the LINAC. The RL signals were then calibrated with respect to reference dose and dose rate data using an ionization chamber (IC). The algorithm relies upon the integral of the RL and provides the accumulated dose (useful to the medical physicist) at any time during irradiation, the dose rate being derived afterwards. It is tested with both step and arbitrary dose rate profiles, manually operated from the LINAC control desk. The doses measured by RL and OSL are both compared to reference doses and deviations are about ±2% and ±1% respectively, thus demonstrating the reliability of the algorithm for arbitrary profiles and wide range of dose rates. Although the calculation was done off-line, it is amenable to real-time processing during irradiation
Complete list of metadatas

Cited literature [20 references]  Display  Hide  Download

https://hal-cea.archives-ouvertes.fr/cea-01802689
Contributor : Bruno Savelli <>
Submitted on : Thursday, March 19, 2020 - 12:41:29 PM
Last modification on : Thursday, March 19, 2020 - 5:12:55 PM
Long-term archiving on: : Saturday, June 20, 2020 - 2:57:51 PM

File

article_SylvainMagne_IEEE-TNS_...
Files produced by the author(s)

Identifiers

Collections

CEA | DRT | LIST | DM2I

Citation

Sylvain Magne, Sybelle Deloule, Aimé Ostrowsky, Pierre Ferdinand. Fiber-Coupled, Time-Gated Al$_2$O$_3$:C Radioluminescence Dosimetry Technique and Algorithm for Radiation Therapy With LINACs. IEEE Transactions on Nuclear Science, Institute of Electrical and Electronics Engineers, 2013, 60 (4), pp.2998 - 3007. ⟨10.1109/TNS.2013.2263640⟩. ⟨cea-01802689⟩

Share

Metrics

Record views

289

Files downloads

691