Magnetocardiography measurements with $^4$He vector optically pumped magnetometers at room temperature

Abstract : In this paper, we present a proof of concept study which demonstrates for the first time the possibility to record magnetocardiography (MCG) signals with $^4$He vector optically-pumped magnetometers (OPM) operated in a gradiometer mode. Resulting from a compromise between sensitivity, size and operability in a clinical environment, the developed magnetometers are based on the parametric resonance of helium in zero magnetic field. Sensors are operated at room-temperature and provide a tri-axis vector measurement of the magnetic field. Measured sensitivity is around 210 fT/√Hz in the bandwidth [2 Hz; 300 Hz]. MCG signals from a phantom and two healthy subjects are successfully recorded. Human MCG data obtained with the OPMs are compared to reference electrocardiogram (ECG) recordings: similar heart rates, shapes of the main patterns of the cardiac cycle (P/T waves, QRS complex) and QRS widths are obtained with both techniques.
Document type :
Journal articles
Complete list of metadatas

https://hal-cea.archives-ouvertes.fr/cea-01560997
Contributor : Etienne Labyt <>
Submitted on : Wednesday, July 12, 2017 - 11:29:28 AM
Last modification on : Thursday, August 6, 2020 - 3:00:47 AM
Long-term archiving on: : Wednesday, January 24, 2018 - 9:33:30 PM

File

MCG_He4magnetometer_2017_Phys....
Publication funded by an institution

Identifiers

Collections

CEA | CNRS | DRT | UGA | LETI | G2ELAB | CEA-GRE

Citation

S. Morales, M.C. Corsi, W. Fourcault, F. Bertrand, G. Cauffet, et al.. Magnetocardiography measurements with $^4$He vector optically pumped magnetometers at room temperature . Physics in Medicine and Biology, IOP Publishing, 2017, ⟨10.1088/1361-6560/aa6459⟩. ⟨cea-01560997⟩

Share

Metrics

Record views

457

Files downloads

1867